Spider Pin - C1 - C2 Placement
Follow points 1 to 4 as above.

- After site disinfection (chlorhexidine) the spiral drill is used to perforate
the soft tissue and cortical bone (no incision needed). Cold irrigation is
used during the drilling procedure (5°C/41°F). Use the 0.9 mm drill for
the PIN, 1.2 mm drill for the C1 and 1.5 mm for C2 Spider Screws.
- You can choose between two options: manual or mechanical insertion.
For manual insertion use DSX-1690S + DSP-2352S for PIN, DSX-1690S
+ DSP-5052S for C1 and DSX-1690S + DSP-6052S for C2 Spider
Screws. For mechanical insertion use the contra angle pick-up driver
DPQ-2322 for PIN, DPQ-2820 for C1 or DPQ-4422 for C2 Spider
Screws mounted on a low speed contra-angle handpiece (25/30 rpm).
Final placement is achieved by using the handle driver to complete
the insertion as this provides the most controlled tactile method. 6a