Spider Pin - C1 - C2 Placement

Follow points 1 to 4 as above.

  1. After site disinfection (chlorhexidine) the spiral drill is used to perforate the soft tissue and cortical bone (no incision needed). Cold irrigation is used during the drilling procedure (5°C/41°F). Use the 0.9 mm drill for the PIN, 1.2 mm drill for the C1 and 1.5 mm for C2 Spider Screws.
  2. You can choose between two options: manual or mechanical insertion. For manual insertion use DSX-1690S + DSP-2352S for PIN, DSX-1690S + DSP-5052S for C1 and DSX-1690S + DSP-6052S for C2 Spider Screws. For mechanical insertion use the contra angle pick-up driver DPQ-2322 for PIN, DPQ-2820 for C1 or DPQ-4422 for C2 Spider Screws mounted on a low speed contra-angle handpiece (25/30 rpm). Final placement is achieved by using the handle driver to complete the insertion as this provides the most controlled tactile method. 6a